| pH Range | -2.00 to 16.00 pH |
| Resolution & Accuracy | 0.01 pH & ±0.01 pH |
| mV Range | -1000 to 1000 mV / 0 to 100% (for -100 to 100 mV) |
| Resolution & Accuracy | 1 mV / 0.1% & ±1 mV / ±0.2% |
| Temperature Range | -10.0 to 110 °C (14.0 to 230.0 °F) |
| Resolution & Relative Accuracy | 0.1 °C / °F & ±0.5 °C / ±1.0 °F |
| Temperature Sensor | Pt100/ Pt1000 (jumper selectable); 2 or 3 wire |
| Temperature Compensation | Auto / Manual (Reference at 25°C) |
| Set Point and Controller Functions |
| Integral Time | 0 to 999.9 minutes |
| Switching pH Hysteresis | 0.1 to 1 pH |
| Function (Switchable) | P/PI control (Pulse Length / Pulse Frequency); Limit Control; Off |
| Adjustable Period with Pulse Length Controller | 0.5 to 20 seconds |
Adjustable Period with Pulse Frequency Controller
| 60 to 120 pulses/min |
| Pickup / Dropout Delay | 0 to 2000 seconds |
| Wash Cycle | 0.1 to 200.0 hours |
| Wash Duration | 1 to 2000 seconds |
Contact Outputs, Controller
| 1 SPDT, 3 SPST relays |
| Switching Voltage / Current / Power | Max 250 VAC / Max 3A / Max 600VA |
| Alarm Functions |
Function (Switchable)
| Latching / pulse |
| Pickup Delay | 0 to 2000 seconds |
| Display |
LCD
| UV coat, backlit 14 segments display with symbols for status information |
| Backlight | On/Off selectable with four levels of brightness control |
| Electromagnetic Compliance (EMC) Specifications |
Emitted Interference | According to EN 61326 |
| Immunity to Interference | According to EN 61326 |
Environmental Conditions |
| Ambient Temperature Operating Range | 0 to 40 °C |
| Maximum Relative Humidity | 80% up to 31 °C decreasing linearly to 50% at 40 °C |
| Power Supply |
Input
| 80 to 250 VAC/DC 50/60 Hz Approx. 10 VA |
Main Fuse | 250 mA, Anti-surge, S504 BUSSMAN |
| Pollution Degree | 2 |
| Transient Overvoltage Category | II |
| Electrical Data and Connections |
| Transmitter Function | Two 0/4 to 20 mA scalable outputs for pH/ORP and temperature, galvanically isolated |
| CU 22 Function | 22 mA current output |
| Isolated Output Voltage | +12V ±0.5 V (Max 50 mA) |
Hold Function Switch
| To freeze output current and deactivate control relays |
| Load | Max 600 W |
| pH Input | BNC (1012 impedance); Asymmetrical/ Symmetrical |
Connection Terminal (Wall Mount) | 3-pin, 8-pin, 9-pin and 13-pin terminal, detachable blocks |
Connection Terminal (Panel Mount) | 3-pin, 9-pin and 19-pin terminal, detachable blocks |
| Mechanical Specifications |
| Dimensions (W x H x D) | Wall-mount Panel-mount | 144 x 144 x 110 mm 96 X 96 X 175 mm |
| Weight | Wall-mount Panel-mount | 745g (Unit) / 1100g (Packed) 950g (Unit) / 1100g (Packed) |
| Environmental Rating | Wall-mount Panel-mount | IP66 (NEMA 4X) IP54 (front panel) |